Type: Task
-
Status: Open (View Workflow)
-
Priority: Normal
-
Resolution: Unresolved
-
Component/s: ics_xcuActor
-
Labels:None
-
Story Points:1
It turns out that there is a semi-documented (not in table, but in narrative text) register for bth reading and writing whether the 4UHV ionpump controller. Register 8, 0/1/2 for serial/remote/local. We want serial for operations. And yes, you can remotely change from local to serial. Which may be why they do not document it well....