SM1 at Subaru : shutters failed

    XMLWordPrintable

    Details

    • Story Points:
      2
    • Sprint:
      SM1PD-2020 C, SM1PD-2020 D, SM1PD-2020 G, SM1PD-2020 G2, SM1PD-2020 H, SM1PD-2020 J, SM1PD-2021 A 3, SM1PD-2021 A 4, SM1PD-2021 A 5, SM1PD-2021 A 6, SM1PD-2021 A 11

      Description

       

      03:53:07.430948 spsait w text="enu_sm1 shutters failed: FysomError('event init inappropriate in current state FAILED',) in fn() at /software/conda/lib/python3.6/site-packages/fysom/_init_.py:272"

      Don't know why they failed

      Basically biasha board is stuck in a peculiar state where it doesn't answer correctly.

      alefur@shell-ics:~$ oneCmd.py enu_sm1 biasha raw=status                                                            
      2020-08-13T16:00:44.454 sent enu_sm1 biasha raw=status (from alefur_shell_ics)                                     
      2020-08-13T16:00:44.478 enu_sm1 : text="returned: 30"                                                              
      alefur@shell-ics:~$ oneCmd.py enu_sm1 biasha raw=statword                                                          
      2020-08-13T16:01:01.772 sent enu_sm1 biasha raw=statword (from alefur_shell_ics)                                   
      2020-08-13T16:01:01.792 enu_sm1 : text="returned: n"
      

       The only fix is to power cycle the board