"gatevalve open" should check pump status, not just pressures.

    Details

      Description

      The gatevalve open command from INSTRM-406 checks cryostat and roughing line pressures before opening. This is sufficient to be safe, but we should also check both turbo and roughing pump status just to be complete.

        Attachments

          Issue Links

            Activity

            Hide
            cloomis cloomis added a comment -

            Merged at 92c3a98, tagged 1.7.12

            Only checking pump speeds, but I think that is enough.

            To open the gate valve underVacuum, both pumps must be running at the slow-to-reach full speed, which may annoy some.

            Show
            cloomis cloomis added a comment - Merged at 92c3a98, tagged 1.7.12 Only checking pump speeds , but I think that is enough. To open the gate valve underVacuum, both pumps must be running at the slow-to-reach full speed, which may annoy some.