Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2608

Update saturation level for AG cameras

    XMLWordPrintable

    Details

      Description

      With INSTRM-2558, saturation level is now handled each amp separately per region. A few amps are know to show saturation lower than 65535. In the run 23 in June 2025, a couple of data set were taken to analyze the saturation level.

      Note that once saturation level is updated, "flat-top" flag will be removed (INSTRM-2586).

       

      The visit=127134 and 127480 are the data.

      As noted in INSTRM-2558, Kawanomoto-san kindly looked at the data visit=127134 separately:

      • AG2 Left looks saturated around 64200
      • AG3 Left looks saturated around 60500
      • There doesn't seem bright enough object on AG2 Right and AG4 Left.