Type:
Bug
-
Status: Done (View Workflow)
-
Priority:
Normal
-
Resolution: Done
-
Component/s: ics_utils
-
Labels:
-
Story Points:1
-
Sprint:PreRun22May2025, PreRun23Jun2025
INSTRM-2200 introduced a robust way of latching the red resolution but left a bug behind
If you move the grating but don't take a shutter exposure, redResolution keyword is not being updated and as a result, some bias/dark can end up a wrong W_ARM value even though the filename is correct.
It's bad because that causes the ingestion to fail in the drpActor.
So I guess that show two problem :
- two different ways to detect the arm, one say low the other say med, that should be unified.
- redResolution has a visit field, that should be checked against the current visit
125108 is one of those visits :
(rubin8-ics) [alefur@shell-ics sps]$ fitsheader PFSA12510812.fits |grep 'W_ARM' W_ARM = 4 / Spectrograph arm 1=b, 2=r, 3=n, 4=medRed