Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2339

Update FPA position of all visible cameras after adjusting cooler set temperature

    XMLWordPrintable

    Details

    • Sprint:
      preRun18August1

      Description

      Because the cryocooler of all visible cameras didn't work in temperature mode correctly, set temperature was adjusted. The camera temperature and hence focus position changed accordingly. (Focus shifted by up to ~20um.) The new position is

      • xcu_r1 motors moveCcd a=160.65 b=136.61 c=114.20 microns abs
      • xcu_r2 motors moveCcd a=165.22 b=193.73 c=172.15 microns abs
      • xcu_r3 motors moveCcd a=206.96 b=197.48 c=197.04 microns abs
      • xcu_r4 motors moveCcd a=118.70 b=71.72 c=113.78 microns abs
      • xcu_b1 motors moveCcd a=141.88 b=115.56 c=87.96 microns abs
      • xcu_b2 motors moveCcd a=228.86 b=240.71 c=230.06 microns abs
      • xcu_b3 motors moveCcd a=192.85 b=176.70 c=194.74 microns abs
      • xcu_b4 motors moveCcd a=136.00 b=103.91 c=149.77 microns abs