Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2275

Change meb alert threshold and target

    XMLWordPrintable

    Details

      Description

      During the Run 16 (2-8 May 2024), we noticed alert from meb Actor, claiming that temperature of "MEB Top Plate" is out of range (< 0degC). From conversation at that time. we agreed to remove the lower threshold of the alert. Also, I think we should change the monitoring target to proper one because "top plate" is namely the top plate of MCS structure, close to the interface plate. To monitor electronics, "MEB Electronics Rack Temp" should be proper

        Attachments

          Activity

            People

            • Assignee:
              Unassigned
              Reporter:
              yuki.moritani yuki.moritani
            • Votes:
              0 Vote for this issue
              Watchers:
              1 Start watching this issue

              Dates

              • Created:
                Updated: