Type: Task
-
Status: Won't Fix (View Workflow)
-
Priority: Normal
-
Resolution: Won't Fix
-
Component/s: ics_xcuActor
-
Labels:None
When SpS cameras were recovered from maintenance power outage on 3/20, some modules were off by default for r4 and b4 camera. INSTRM-2066 adjusted for SM1-3 cameras, but was not applied to r4/b4 cameas.
(rubin3_ics) moritani@shell-ics:~$ oneCmd.py xcu_r3 pcm getMask powerOn 2024-03-23T22:22:17.723 sent xcu_r3 pcm getMask powerOn (from moritani_shell_ics) 2024-03-23T22:22:17.742 xcu_r3 : text="returned b'01011011'" (rubin3_ics) moritani@shell-ics:~$ oneCmd.py xcu_r4 pcm getMask powerOn 2024-03-23T22:22:08.637 sent xcu_r4 pcm getMask powerOn (from moritani_shell_ics) 2024-03-23T22:22:10.078 xcu_r4 : text="returned b'00010000'" (rubin3_ics) moritani@shell-ics:~$ oneCmd.py xcu_b4 pcm getMask powerOn 2024-03-23T22:22:25.254 sent xcu_b4 pcm getMask powerOn (from moritani_shell_ics) 2024-03-23T22:22:25.274 xcu_b4 : text="returned b'00010000'"
The same mask should be set as the other cameras.