Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2013

reported readTimes incorrect

    XMLWordPrintable

    Details

      Description

      The hxActor reports the expected times for a ramp as soon as it knows. That includes the time for a single read (which should be calculated from the ASIC times). That is somehow getting changed/miscalculated:

      2023-07-06 07:18:14.245Z cmds             20 Actor.py:524 new cmd: ramp nread=113 visit=96389 exptype=object expectedExptime=1200.0
      2023-07-06 07:18:18.760Z cmds             20 CommandLink.py:122 > 2 444 i readTimes=96389,2023-07-05T21:18:18.749665,2023-07-05T21:18:29.606715,10.857
      2023-07-06 07:38:46.012Z cmds             20 Actor.py:524 new cmd: ramp finish exptime=1200.091 obstime=2023-07-05T21:18:42.394797 stopRamp
      2023-07-06 07:38:57.717Z cmds             20 CommandLink.py:122 > 2 444 i readTime=42.3220
      2023-07-06 07:39:23.960Z cmds             20 Actor.py:524 new cmd: ramp nread=31 visit=96390 exptype=object expectedExptime=1200.0
      2023-07-06 07:39:27.987Z cmds             20 CommandLink.py:122 > 2 446 i readTimes=96390,2023-07-05T21:39:27.978713,2023-07-05T21:40:10.300723,42.322
      

        Attachments

          Issue Links

            Activity

              People

              • Assignee:
                cloomis cloomis
                Reporter:
                cloomis cloomis
              • Votes:
                0 Vote for this issue
                Watchers:
                1 Start watching this issue

                Dates

                • Created:
                  Updated: