<item>
            <title>[INSTRM-2007] Add changes for new HgCd lamp control.</title>
                <link>https://pfspipe.ipmu.jp/jira/browse/INSTRM-2007</link>
                <project id="10300" key="INSTRM">Instrument control development</project>
                    <description>&lt;p&gt;The new HgCd hardware has been installed, and the basic software configuration tested. The changes need to be added to the `calibhead` and `caliblamps` files in the pfilampsActor product.&lt;/p&gt;</description>
                <environment></environment>
        <key id="23653">INSTRM-2007</key>
            <summary>Add changes for new HgCd lamp control.</summary>
                <type id="3" iconUrl="https://pfspipe.ipmu.jp/jira/secure/viewavatar?size=xsmall&amp;avatarId=10518&amp;avatarType=issuetype">Task</type>
                                            <priority id="10000" iconUrl="https://pfspipe.ipmu.jp/jira/images/icons/priorities/medium.svg">Normal</priority>
                        <status id="10002" iconUrl="https://pfspipe.ipmu.jp/jira/images/icons/statuses/generic.png" description="The issue is resolved, reviewed, and merged">Done</status>
                    <statusCategory id="3" key="done" colorName="green"/>
                                    <resolution id="10000">Done</resolution>
                                        <assignee username="cloomis">cloomis</assignee>
                                    <reporter username="cloomis">cloomis</reporter>
                        <labels>
                            <label>near-term</label>
                    </labels>
                <created>Wed, 5 Jul 2023 15:44:35 +0000</created>
                <updated>Fri, 4 Aug 2023 12:47:18 +0000</updated>
                            <resolved>Fri, 14 Jul 2023 14:01:24 +0000</resolved>
                                                                    <component>ics_pfilampsActor</component>
                        <due></due>
                            <votes>0</votes>
                                    <watches>1</watches>
                                                                <comments>
                            <comment id="33990" author="cloomis" created="Mon, 10 Jul 2023 22:15:50 +0000"  >&lt;p&gt;We have been testing the new lamps. The changes are mostly configuration of the photodiode signal to require before turning off the HV, and of the fan level to set after that. It looks like we will settle on a warmup period close to the hardware limit of three minutes &amp;#8211; that nearly doubles the measured Cd flux over two minutes. Still working on the fan level, but we suspect that will need to be chosen based on what we see on the detectors with the lamps on the PFI. Might need to be adaptive.&lt;/p&gt;</comment>
                            <comment id="34017" author="cloomis" created="Fri, 14 Jul 2023 14:01:24 +0000"  >&lt;p&gt;The best performance as tested off the telescope comes from warming up to a threshold of Cd=0.6 (in internal units), which takes the full 180s before the hardware cutoff. Then running the fan at 15%. Warmups take about the same amount of time if done again after powering the lamp down, so operationally I&apos;m guessing that shutter-controlled illumination is the best choice.&lt;/p&gt;

&lt;p&gt;If we need to make the Cd lines brighter it should be possible by adding active fan control. As it stands if the fan is run too slowly (~10%) the flux increases uncomfortably after a few minutes.&lt;/p&gt;</comment>
                            <comment id="34249" author="cloomis" created="Fri, 4 Aug 2023 12:47:18 +0000"  >&lt;p&gt;Tagged 1.3.2&lt;/p&gt;</comment>
                    </comments>
                    <attachments>
                    </attachments>
                <subtasks>
                    </subtasks>
                <customfields>
                                                <customfield id="customfield_10500" key="com.atlassian.jira.plugins.jira-development-integration-plugin:devsummary">
                        <customfieldname>Development</customfieldname>
                        <customfieldvalues>
                            
                        </customfieldvalues>
                    </customfield>
                                                                                                                                                                                                            <customfield id="customfield_10010" key="com.pyxis.greenhopper.jira:gh-lexo-rank">
                        <customfieldname>Rank</customfieldname>
                        <customfieldvalues>
                            <customfieldvalue>0|02qpjd:00r20060i200186403r7i</customfieldvalue>

                        </customfieldvalues>
                    </customfield>
                                                                                                                                                                                                                                        </customfields>
    </item>
</channel>