[PIPE2D-879] Repeat data acquisition for DCB in blue with HgAr Created: 05/Aug/21 Updated: 02/Oct/21 Resolved: 02/Oct/21 |
|
| Status: | Done |
| Project: | DRP 2-D Pipeline |
| Component/s: | None |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Task | Priority: | Normal |
| Reporter: | ncaplar | Assignee: | ncaplar |
| Resolution: | Done | Votes: | 0 |
| Labels: | None | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Issue Links: |
|
||||||||
| Story Points: | 2 | ||||||||
| Description |
|
While taking data with HgAr lamp during June/July acquisition I made a mistake and I took the data in the same fashion as with noble gas lamps. Given that the flux of HgAr lamp is a nonlinear function of time, I have seriously overexposed the defocused data, and achieved very low flux in the focused data. This data acquisition has to be repeat, either by using a shutter or by carefully modifying exposure times. During SPS telecon on Wen, August 4 2021, it has been projected that the possibility for this work might exist during end of August, beginning of September. |
| Comments |
| Comment by ncaplar [ 05/Aug/21 ] |
|
Part of this analysis could be done with available Neon data. |
| Comment by ncaplar [ 08/Sep/21 ] |
|
As discussed in #commisioning, it might be possible to take some data at the end of this week (which started on September 6) |