[PIPE2D-729] SuNSS biases show lines Created: 17/Feb/21  Updated: 05/Mar/21  Resolved: 05/Mar/21

Status: Won't Fix
Project: DRP 2-D Pipeline
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Story Priority: Normal
Reporter: hassan Assignee: cloomis
Resolution: Won't Fix Votes: 0
Labels: SuNSS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File Screen Shot 2021-02-16 at 8.15.41 PM.png     PNG File Screen Shot 2021-02-16 at 9.13.10 PM.png    
Story Points: 2
Sprint: 2DDRP-2021 A 2, 2DDRP-2021 A3
Reviewers: hassan

 Description   

Recent bias exposures taken from SuNSS show lines that appear for all fibers. See for example

This may be due to incorrect wiping of the CCDs, eg by not running the parallel clocks. The actual source of these lines may be work/room lamps being left on during data acquisition.

Find the cause for the lines and fix.



 Comments   
Comment by rhl [ 17/Feb/21 ]

Ummh, that's a 10s exposure, so it's not mysterious that there should be spectral features from Hg lamps in the dome.  The weird thing is the r-band biases such as 45775 which also show what look like lines, despite the fact that there should be no time when we're not running the parallels.

Comment by hassan [ 17/Feb/21 ]

Yes, indeed sorry - here is the visit=45775 r-band image :

Comment by rhl [ 05/Mar/21 ]

These are real Hg lines. There are c. 3s of "staring" time during the readout (cf. PIPE2D-743), and that's enough to pick up lines from the dome lights.