| Description |
|
Measure the positional difference of dome flat data taken at 4 different ring lamp positions. |
| Comments |
| Comment by Wilfred Gee [ 30/Jan/24 ] | ||||||||||||||||||||||||
|
Relevant visits:
| ||||||||||||||||||||||||
| Comment by Wilfred Gee [ 30/Jan/24 ] | ||||||||||||||||||||||||
|
The two box plots (one each for `r` and `b`) show the iqr for the median `dx` pixel position of each fiber as well as any significant outliers. The color coding on the row background corresponds to the different ring lamps.
There appears to be no positional difference whatsoever for the different dome flats taken with the various ring lamps. This is reflected in the generated 2D maps as well, which don't show any visual difference.
`r4` has a larger `dx` compared to the other `r` spectrographs but they are still consistent with each other.
| ||||||||||||||||||||||||
| Comment by Wilfred Gee [ 30/Jan/24 ] | ||||||||||||||||||||||||
|
I've attached the 2D `dx` detector map residuals for visit 104640 for each detector, which are representative of all the dome visit residuals. |