[INSTRM-926] Perform ISR on AG image Created: 20/Mar/20  Updated: 28/Jul/22  Resolved: 28/Jul/22

Status: Won't Fix
Project: Instrument control development
Component/s: ics_agccActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: hassan Assignee: chyan
Resolution: Won't Fix Votes: 0
Labels: EngRun
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: Text File agc_dark_exposures.txt    
Issue Links:
Blocks
blocks INSTRM-929 Perform object detection and centroid... Done
is blocked by INSTRM-1469 Create a Detector model for the AG CCDs Done
Relates
relates to INSTRM-1474 Correctly process raw images before c... Done
Story Points: 2
Sprint: PreEngRun05 A, PreEngRun05 B, PreEngRun05 C, PreEngRun05 D, PreEngRun05 E, PreEngRun05 F

 Description   

Apply instrument signature removal (ISR) on the supplied AG image. This is a necessary step prior to centroiding.



 Comments   
Comment by rhl [ 21/Mar/20 ]

I don't know how much ISR (if any) will be required for these devices, but we should keep the possibility open for now

Comment by hassan [ 27/Mar/20 ]

The ISR should complete in less than 1s in order to keep within the field acquisition budget.

Comment by hassan [ 03/Dec/21 ]

Flat data has been generated, which ASIAA can use.

Comment by rhl [ 03/Dec/21 ]

Have flats been generated, or just raw flats taken? I have a Detector model for the AG cameras, which allows us to use standard Rubin/PFS code to make master calibrations. I do need to check it in...

Comment by yuki.moritani [ 03/Dec/21 ]

rhl I sent the list of the flat during the run. If you were happy with them, ASIAA can create the flat using them.

The flat image list is:

2021-09-24/agcc_20210924_0156267.fits
2021-09-24/agcc_20210924_0156291.fits
2021-09-24/agcc_20210924_0156317.fits
2021-09-24/agcc_20210924_0156343.fits
2021-09-24/agcc_20210924_0156368.fits
2021-09-24/agcc_20210924_0156443.fits
2021-09-24/agcc_20210924_0156469.fits
2021-09-24/agcc_20210924_0156495.fits
2021-09-24/agcc_20210924_0156520.fits
2021-09-24/agcc_20210924_0156547.fits
Comment by chyan [ 03/Dec/21 ]

I was about to ask yuki.moritani the list of the flats and darks. rhl, may I also take a look of your Detector model code.  I can think about how to put them into agccActor.  Thanks.

Comment by yuki.moritani [ 03/Dec/21 ]

And "agc_dark_exposures.txt" lists dark frames... although one frame per each were taken... so we may need to take another data.

Comment by hassan [ 21/Dec/21 ]

rhl could you contact Chi-Hung regarding your detector model code? Has it been checked in yet?

Comment by hassan [ 11/Mar/22 ]

Chi-Hung expect to address this by Mar 14.

Comment by chyan [ 18/Mar/22 ]

Starting to work on this issue.

Comment by hassan [ 15/Apr/22 ]

Following clarifications during the ICS/PFI telecon 2022-04-15: Princeton will provide the calibration products for ASIAA to proceed. rhl will speak to Paul about this.

Comment by chyan [ 28/Jul/22 ]

I closing this ticket simply because AG performance is verified OK without ISR.

Generated at Sat Feb 10 16:30:04 JST 2024 using Jira 8.3.4#803005-sha1:1f96e09b3c60279a408a2ae47be3c745f571388b.