[INSTRM-924] Add command to control local/remote/serial ionpump modes Created: 18/Mar/20 Updated: 18/Mar/20 |
|
| Status: | Open |
| Project: | Instrument control development |
| Component/s: | ics_xcuActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Task | Priority: | Normal |
| Reporter: | cloomis | Assignee: | cloomis |
| Resolution: | Unresolved | Votes: | 0 |
| Labels: | None | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Story Points: | 1 |
| Description |
|
It turns out that there is a semi-documented (not in table, but in narrative text) register for bth reading and writing whether the 4UHV ionpump controller. Register 8, 0/1/2 for serial/remote/local. We want serial for operations. And yes, you can remotely change from local to serial. Which may be why they do not document it well.... |