[INSTRM-893] STS alarm message/email does not forward text field Created: 06/Feb/20 Updated: 14/Oct/20 Resolved: 08/Oct/20 |
|
| Status: | Done |
| Project: | Instrument control development |
| Component/s: | None |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Task | Priority: | Normal |
| Reporter: | arnaud.lefur | Assignee: | Yoshida, Hiroshige |
| Resolution: | Done | Votes: | 0 |
| Labels: | SPS | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Issue Links: |
|
||||||||
| Sprint: | SM1PD-2020 J | ||||||||
| Description |
|
STS alarm messages that we receive contain only the value field : PFS: XCU_R1 Cooler1 Load value is at 0 ACK27932 Mon Jan 13 04:36:08 2020 But text field has information in it that help operator to diagnose the problem, so we definitely need it. Is that trivial to add Yoshida, Hiroshige ? |
| Comments |
| Comment by Yoshida, Hiroshige [ 06/Feb/20 ] |
|
Add a dollar sign character "$" somewhere in your Alarm Message. It will be replaced with a text value. An asterisk character "*" will be replaced with a numeric value. For example, the full message for "PFS: XCU_R1_Cooler1_Load" is currently:
PFS: XCU_R1_Cooler1_Load value is *
Change it to something like:
PFS: XCU_R1_Cooler1_Load value is *, $
|
| Comment by hassan [ 08/Oct/20 ] |
|
Re-assigned to Yoshida, Hiroshige to implement the corresponding changes in the STS. cloomis and arnaud.lefur in discussions today agree with his proposal and will address that in |
| Comment by Yoshida, Hiroshige [ 08/Oct/20 ] |
|
Except the ion pump pressures, it seems that configuration changes for the SM1 data source alarms have already been applied by a STS user. |
| Comment by hassan [ 08/Oct/20 ] |
|
So Yoshida, Hiroshige would you be able to arrange for the corresponding configuration changes for ion pump pressures, then we can close this ticket? |
| Comment by arnaud.lefur [ 08/Oct/20 ] |
|
I think there is a small misunderstanding, this ticket should be closed, Yoshida, Hiroshige gave us a solution, that will be soon improved ( |
| Comment by Yoshida, Hiroshige [ 08/Oct/20 ] |
|
I added the text data to the ionpump pressure alarms. |
| Comment by hassan [ 08/Oct/20 ] |
|
Thanks Yoshida, Hiroshige for your updates. |