[INSTRM-873] Test if sm2 red exchange mechanism behaviour is similar to sm1 one Created: 08/Jan/20  Updated: 23/Jan/20  Resolved: 17/Jan/20

Status: Done
Project: Instrument control development
Component/s: ics_enuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: arnaud.lefur Assignee: arnaud.lefur
Resolution: Done Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Story Points: 1
Sprint: SM1PD-2019 E, SM1PD-2020 A

 Description   

SM1 red exchange mechanism is showing some problem since holding current has indeed been deactivated : after being set in position, either low or medium resolution, limit switch signal turns to 0 after some time.

Since none of the limit switches are activated, rexm is therefore in anĀ unkown position.

SM2 rexm is now operational at LAM, i'm testing it to see if this behaviour is reproducible.



 Comments   
Comment by arnaud.lefur [ 17/Jan/20 ]

Problem is indeed reproducible, winlight is now performing some test on sm3