[INSTRM-80] Ion pump start error: only one started Created: 17/Jan/17  Updated: 17/Jun/17  Resolved: 17/Jun/17

Status: Done
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Minor
Reporter: fmadec Assignee: cloomis
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: Text File LOCAL.txt    
Issue Links:
Duplicate
duplicates INSTRM-81 ion pump number 2 start is long and b... Done

 Description   

10:25:48.617716 .hub 0 cmds d CmdIn="client.pfs","xcu_r1","ionpump on"
10:25:48.618379 .hub 0 cmds d CmdQueued=1012496,1484648748.60,"client.pfs",16,"xcu_r1",90,"ionpump on"
10:25:49.385994 client.pfs 16 xcu_r1 i ionPump1=1,300,2.6e-06,35,1.2e-08
10:25:50.071941 client.pfs 16 xcu_r1 i ionPump2=0,0,0,35,0
10:25:50.072684 client.pfs 16 xcu_r1 f text="ion pump controller is in LOCAL mode!"
10:25:50.073346 .hub 0 cmds d CmdDone=1012496,"f"

10:26:26.672752 .hub 0 cmds d CmdIn="client.pfs","xcu_r1","ionpump status"
10:26:26.673473 .hub 0 cmds d CmdQueued=1012539,1484648786.67,"client.pfs",17,"xcu_r1",91,"ionpump status"
10:26:27.419803 client.pfs 17 xcu_r1 i ionPump1=1,6100,0.013,35,8.5e-05
10:26:28.146249 client.pfs 17 xcu_r1 i ionPump2=0,0,0,35,0
10:26:28.146943 client.pfs 17 xcu_r1 :
10:26:28.147526 .hub 0 cmds d CmdDone=1012539,":"



 Comments   
Comment by fmadec [ 17/Jan/17 ]

after a start again, it works

10:32:32.453227 .hub 0 cmds d CmdIn="client.pfs","xcu_r1","ionpump on"
10:32:32.453983 .hub 0 cmds d CmdQueued=1012940,1484649152.45,"client.pfs",18,"xcu_r1",92,"ionpump on"
10:32:33.240493 client.pfs 18 xcu_r1 i ionPump1=1,7000,0.00036,35,5.9e-07
10:32:34.107678 client.pfs 18 xcu_r1 i ionPump2=1,0,1.7e-06,33,8.8e-09
10:32:34.854146 client.pfs 18 xcu_r1 i ionPump1=1,7000,0.00036,35,5.9e-07
10:32:35.600488 client.pfs 18 xcu_r1 i ionPump2=1,900,1.3e-06,33,7e-09

Comment by cloomis [ 17/Jan/17 ]

The reported error:
10:25:50.072684 client.pfs 16 xcu_r1 f text="ion pump controller is in LOCAL mode!"

did indeed come back from command to turn pump 2 on. Specifically, it is "The window specified is Read Only or is temporarily disabled.", but we did see it for commands sent in local mode.

Given the slightly different problem in INSTRM-74, I'm suspect we have to be gentler about commanding this thing.

Comment by cloomis [ 17/Jan/17 ]

And INSTRM-81. Are we having RS-485 problems?

Comment by cloomis [ 17/Jun/17 ]

Finished at 00d432c