[INSTRM-771] Shutdown cryocooler when reject temperature threshold hit. Created: 18/Sep/19  Updated: 03/Oct/19  Resolved: 24/Sep/19

Status: Done
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: SM1
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Story Points: 1
Sprint: SM1-2019 N, SM1-2019 O
Reviewers: arnaud.lefur

 Description   

If status shows reject temperature above some threshold temperature, shutdown the cooler loop. Also generate a synthetic error bit & message for the alertsActor to act on.



 Comments   
Comment by cloomis [ 24/Sep/19 ]
  • status command stops cooler loop if reject temperature above limit.
  • Added synthetic error bit (0x100) to indicate this condition
  • limit currently set in .cfg file.
Comment by cloomis [ 24/Sep/19 ]

Tested on b2.