[INSTRM-758] Add synthetic error bit for INSTRM-594, append description field to ionpumpNerrors Created: 05/Sep/19  Updated: 05/Sep/19  Resolved: 05/Sep/19

Status: Done
Project: Instrument control development
Component/s: ics_actorkeys, ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Issue Links:
Relates
relates to INSTRM-594 Alert on voltage/current anomalies in... Done
Story Points: 1

 Description   

See INSTRM-594. Add 2^12 bit when pump state is on but current and/or pressure are 0.

Also add string field to ionpumpNerrors key, describing all set error bits.



 Comments   
Comment by cloomis [ 05/Sep/19 ]

Merged at 4cb50be, tagged 1.10.8.

Tested at JHU. Did not add a string field since it already exists, duh. But did add fleshed-out descriptions for all the Agilent bits and added the synthetic 2^16 as "Suspect live channel", for lack of any better description. arnaud.lefur