[INSTRM-755] Add safety band heater process to SMB. Created: 04/Sep/19  Updated: 20/Sep/19  Resolved: 20/Sep/19

Status: Won't Fix
Project: Instrument control development
Component/s: ics_xcu_smb
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Won't Fix Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Issue Links:
Duplicate
duplicates INSTRM-693 Add autonomous warmup routine. Done
Story Points: 2

 Description   

The temperature board is responsible for trying to keep the detectors warmer than some other things in the cryostat. We have evidence that safe temperature headroom is passively maintained if all power is cut off, but if there is any power available the SMB should still put in some heat.

We must assume that the network is gone and that both the site power and the instrument UPS have failed, so we should be power-conscious.

The proposed scheme is for the heater control loop to switch to a power-efficient bang-bang mode under some condition, not yet nailed down.



 Comments   
Comment by arnaud.lefur [ 04/Sep/19 ]

It's related to INSTRM-693 right ?

Comment by cloomis [ 04/Sep/19 ]

Yes. Effectively a duplicate, sorry. Will close one and do the work under the other.

Comment by hassan [ 20/Sep/19 ]

Duplicates INSTRM-693.

Generated at Sat Feb 10 16:28:19 JST 2024 using Jira 8.3.4#803005-sha1:1f96e09b3c60279a408a2ae47be3c745f571388b.