[INSTRM-705] gauge pressure error Created: 15/Jul/19 Updated: 10/Jul/21 |
|
Status: | In Progress |
Project: | Instrument control development |
Component/s: | ics_xcuActor |
Affects Version/s: | None |
Fix Version/s: | None |
Type: | Bug | Priority: | Minor |
Reporter: | fmadec | Assignee: | shope |
Resolution: | Unresolved | Votes: | 0 |
Labels: | SPS | ||
Remaining Estimate: | Not Specified | ||
Time Spent: | Not Specified | ||
Original Estimate: | Not Specified |
Story Points: | 2 |
Sprint: | SM1-2019 K |
Description |
During testing, the gauge on B1 failed, that was after having start the ionpumps. the same test on R1 was done without this gauge error, pressure level is about the same in both CU.
2019-07-15 10:16:58 cmdIn=xcu_b1 gauge status 2019-07-15 10:16:59 xcu_b1 f text="command failed: ValueError(\"could not convert string to float: 'BUFFER OVERRUN'\",) in parsePressure() at /software/mhs/products/Linux64/ics_xcuActor/1.10.6/python/xcuActor/Controllers/idgPfeiffer.py:58" |
Comments |
Comment by fmadec [ 15/Jul/19 ] |
2019-07-15 10:35:24.059Z actor 30 ICC.py:57 text="controller fee(fee) could not be created" 2019-07-15 10:35:24.060Z cmds 20 CommandLink.py:121 > 2 10 w controllers=ccd 2019-07-15 10:35:24.061Z cmds 20 CommandLink.py:121 > 2 10 w text="failed to connect controller fee 2019-07-15 10:35:24.062Z cmdr 10 CmdrConnection.py:58 read: client.v594.spsait 10 ccd_r1 w controllers=ccd 2019-07-15 10:35:24.065Z cmds 20 CommandLink.py:121 > 2 10 f text="command failed: RuntimeError('failed to arm for readout)',) in pyFPGA.FPGA.setClockLevels() at fpga/pyFPGA.pyx:127" 2019-07-15 10:35:24.104Z cmdr 10 CmdrConnection.py:58 read: client.v594.spsait 10 ccd_r1 w text="failed to connect controller fee" 2019-07-15 10:35:24.105Z cmdr 10 CmdrConnection.py:58 read: client.v594.spsait 10 ccd_r1 f text="command failed: RuntimeError('failed to arm for readout)',) in pyFPGA.FPGA.setClockLevels() at fpga/pyFPGA.pyx:127" |
Comment by cloomis [ 15/Jul/19 ] |
power-cycling gauge cleared the problem, so the error is probably from the gauge and not the PCM. Steve will look into it; please report if it happens again – I have not seen this before.
|
Comment by cloomis [ 24/Jul/19 ] |
Fuzz-tested controller; BUFFER OVERRUN could not be replicated. Park for more evidence. |
Comment by arnaud.lefur [ 10/Jul/21 ] |
happen again today at LAM : 2021-07-09 15:26:17 xcu_r2 f text="command failed: ValueError(\"could not convert string to float: 'BUFFER OVERRUN'\") in parsePressure() at /software/mhs/products/Linux64/ics_xcuActor/1.12.6/python/xcuActor/Controllers/idgPfeiffer.py:58" again after turning ionpump on |