[INSTRM-688] Read the errors for in individual ionpump channels Created: 29/Jun/19  Updated: 29/Jun/19  Resolved: 29/Jun/19

Status: Done
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Story Points: 1

 Description   

The Agilent 4UHV ion pump controller can return the error masks for individual channels or for their union. For no longer valid reasons I used to read the union only. Change this to read and report the errors for individual channels.



 Comments   
Comment by cloomis [ 29/Jun/19 ]

Merged at d260a3d, tagged 1.10.4