[INSTRM-650] W_AITPHO appears to be in foot-lamberts, not cd/m^2 Created: 26/Apr/19  Updated: 10/Jul/19  Resolved: 10/Jul/19

Status: Done
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: arnaud.lefur
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

From a pipe2d slack conversation, referring to the header from PFLA01635112.fits:

W_AITPHO=               12.444 / cd/m^2 at photodiode                   
HIERARCH W_AIT_SRC_diodeFlux = 42.633 / cd/m^2 at photodiode

I think the W_AITPHO value should be 42.633. That is the card we will be keeping.

Oh, while on this, could the cd/m^2 be put into the units field in the keyword dictionary entry?



 Comments   
Comment by arnaud.lefur [ 29/Apr/19 ]

W_AITPHO is in cd/m^2, I'm already applying the conversion.

The old keyword is now incorrect because it's expecting foot-lamberts, so the conversion is done twice. 

You're right, I'll put the unit in dcb actorkeys.

 

Comment by arnaud.lefur [ 10/Jul/19 ]

W_AITPHO is in cd/m² , unit is in the actorkey now

Generated at Sat Feb 10 16:27:14 JST 2024 using Jira 8.3.4#803005-sha1:1f96e09b3c60279a408a2ae47be3c745f571388b.