[INSTRM-6] CCD bias band at amp start. Created: 23/Sep/16  Updated: 13/Oct/17

Status: Open
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Minor
Reporter: cloomis Assignee: Unassigned
Resolution: Unresolved Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File biasEdge.png    

 Description   

There is a band structure seen in the biases, at the start of each amp's readout. The flux rolls up over ~6 pixels from about -15 ADU, stays about 0.5 ADU high for ~30 pixels, then dips to the nominal amp level.

The structure is essentially identical for all amps.

The structure is not visible in bias subtracted superdarks or superflats.

The attached image is from a superbias for the r1 detectors, taken in the optics lab dewar on 2015-12-22. Frames 7251-7259. The amp geometry has been corrected, so this shows the readout for two amps, with time for both starting in the middle of the image and moving out.