| Description |
|
The PFS guide loop will be getting a new focus measurement with each AGCC readout. Arrange to feed that back to Gen2 so that focus can be adjusted during integrations. eric and cloomis need to work this out. |
| Comments |
| Comment by cloomis [ 07/May/22 ] |
|
Current implementation plan: Yoshida, Hiroshige in the agActor will generate an MHS keyword for each AG iteration: agOffsets=$visit,$agSeqno,$dRA,$dDec,$dRot,$dFocus, in arcseconds and microns, with the unmeasured items "nan". cloomis in gen2Actor will convert those to Gen2 status keywords: PFS.AG.[RA_ERR,DEC_ERR,PA_ERR,FOCUS_ERR] |
| Comment by Yoshida, Hiroshige [ 11/May/22 ] |
|
Status keys for the AG actor (guideOffsets & focusOffsets) added. |
| Comment by cloomis [ 06/Aug/22 ] |
|
Was running during 2022-06 engineering run. |