[INSTRM-565] Run close-gatevalve-on-problem thread when gatevalve is open. Created: 06/Nov/18  Updated: 19/Dec/19  Resolved: 19/Dec/19

Status: Won't Fix
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Won't Fix Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Issue Links:
Relates
relates to INSTRM-566 Add site (and possibly instrument) ai... Open
relates to INSTRM-406 Add safety logic to gatevalve control. Done
relates to INSTRM-198 Implement low-level roughing loop mon... Won't Fix
relates to INSTRM-570 Add python-level interlock interrupt ... Won't Fix
Story Points: 3

 Description   

We are careful about opening the gatevalve. We need to actively pay attention to any relevant anomalies (roughing pump status, roughing line pressure, turbo status, communications, site and instrument air pressure, SM power, interlock interrrupts) and close the valve.



 Comments   
Comment by hassan [ 09/Jan/19 ]

This ticket should ideally be broken down into smaller tickets to ensure that the roughing valve closes for each scenario of concern.

Comment by cloomis [ 22/May/19 ]

The new interlock board will handle most real safety concerns, in particular with pressure rising in the roughing line. So this is no longer a critical process. The exception might be the site air; that is still under discussion.

Comment by cloomis [ 19/Dec/19 ]

The new interlock board handles this well enough. Any further requirements should be raised in other tickets.