[INSTRM-564] "gatevalve open" should check pump status, not just pressures. Created: 05/Nov/18 Updated: 09/Jan/19 Resolved: 20/Dec/18 |
|
Status: | Done |
Project: | Instrument control development |
Component/s: | ics_xcuActor |
Affects Version/s: | None |
Fix Version/s: | None |
Type: | Task | Priority: | Normal |
Reporter: | cloomis | Assignee: | cloomis |
Resolution: | Done | Votes: | 0 |
Labels: | SM1 | ||
Remaining Estimate: | Not Specified | ||
Time Spent: | Not Specified | ||
Original Estimate: | Not Specified |
Issue Links: |
|
||||||||||||
Story Points: | 1 | ||||||||||||
Epic Link: | Red channel perf test |
Description |
The gatevalve open command from |
Comments |
Comment by cloomis [ 20/Dec/18 ] |
Merged at 92c3a98, tagged 1.7.12 Only checking pump speeds, but I think that is enough. To open the gate valve underVacuum, both pumps must be running at the slow-to-reach full speed, which may annoy some. |