[INSTRM-53] [xcu_r1] Powering motors pertubates also gauge's reading Created: 27/Dec/16  Updated: 16/Mar/17  Resolved: 16/Mar/17

Status: Done
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Major
Reporter: arnaud.lefur Assignee: shimono
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File 75626ff0-323a-11e6-8631-48f126ed8e8e.png    

 Description   

This issue was initially filed into JIRA by 'arnaud.lefur' at 02/Feb/16 9:02 AM as ICS-24.
This issue was assigned to 'Steve Hope'.

When the motors are powered on, it made pcm to fail to read gauge value from time to time, in a very cyclic way.

text="sending '~320010074002=?106\r\n'"
08:41:04.846958 .xcu_r1 51 xcu_r1 d text="received 'Error: RD_485'"
08:41:04.848455 .xcu_r1 51 xcu_r1 f text="command failed: RuntimeError('Error reading or writing: Error: RD_485',) at /home/pfs/products/Linux64/ics_xcuActor/1.2.1/python/xcuActor/Controllers/PCM.py:63"

When we power off, everythings back to normal.



 Comments   
Comment by cloomis [ 16/Mar/17 ]

Several changes were made to the motor controller power feed in 2017-03, and noise is way down. Closing this.