[INSTRM-365] Values above 50°C seems to be filtered Created: 17/May/18  Updated: 04/Jul/18  Resolved: 04/Jul/18

Status: Done
Project: Instrument control development
Component/s: ics_sps_engineering_plotData
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: fmadec Assignee: arnaud.lefur
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

Values above 50°C seems to be filtered, while during bakeout cryostat body temperature body is higher than 50°C.

I think If data are filtered we should have a flag to disable it, just to to be more generic, ideally we should be able to set the threshold values via the gui.



 Comments   
Comment by arnaud.lefur [ 04/Jun/18 ]

There is a config file which defines those treshold in

 $ICS_SPS_ENGINEERING_LIB_DATAQUERY_DIR/config/datatype.cfg

It's true that we can reach 50°C during bakeout, so I've just set the value to 70°C.

We should be okay now.

Generated at Sat Feb 10 16:24:15 JST 2024 using Jira 8.3.4#803005-sha1:1f96e09b3c60279a408a2ae47be3c745f571388b.