[INSTRM-2309] b2 mangin and detectorbox temperature looks flipped Created: 11/Jun/24 Updated: 12/Jun/24 |
|
Status: | Open |
Project: | Instrument control development |
Component/s: | None |
Affects Version/s: | None |
Fix Version/s: | None |
Type: | Bug | Priority: | Normal |
Reporter: | yuki.moritani | Assignee: | cloomis |
Resolution: | Unresolved | Votes: | 0 |
Labels: | None | ||
Remaining Estimate: | Not Specified | ||
Time Spent: | Not Specified | ||
Original Estimate: | Not Specified |
Issue Links: |
|
||||||||
Story Points: | 1 |
Description |
Looking xcu_b2__vistemps table on archiver, I think values for mangin, detectorbox, and spiderrod is swapped: i.e., mangin looks spiderrod, spiderrod looks detectorbox, and detecotbox looks mangin. Unnamed: 0 id detectorbox mangin spiderrod spreader frontring detector1 detector2 0 0 4593750715 280.402893 146.137405 153.264404 135.693298 282.777710 162.680496 162.988007 1 1 4593699714 280.388214 146.135696 153.260300 135.694397 282.734711 162.679993 162.986893 2 2 4594019579 280.499908 146.151398 153.283203 135.698502 282.989014 162.680695 162.987595 3 3 4593774040 280.410706 146.138397 153.266495 135.693604 282.793915 162.680496 162.987503 4 4 4593791531 280.415009 146.139008 153.267395 135.693298 282.808411 162.679794 162.987106
|
Comments |
Comment by arnaud.lefur [ 11/Jun/24 ] |
(rubin3_ics) alefur@shell-ics:~$ oneCmd.py xcu_b2 temps status 2024-06-11T12:30:28.868 sent xcu_b2 temps status (from alefur_shell_ics) 2024-06-11T12:30:28.886 xcu_b2 i vistemps=275.4443,143.2888,150.4335,133.0566,277.2478,162.6901,162.9803 2024-06-11T12:30:28.886 xcu_b2 : temps=275.4443,143.2888,150.4335,133.0566,277.2478,400.0000,400.0000,400.0000,400.0000,400.0000,162.6901,162.9803 (rubin3_ics) alefur@shell-ics:~$ oneCmd.py xcu_b1 temps status 2024-06-11T12:30:34.333 sent xcu_b1 temps status (from alefur_shell_ics) 2024-06-11T12:30:35.503 xcu_b1 i vistemps=151.0828,275.6105,143.0807,134.9645,277.6646,162.0839,162.9854 2024-06-11T12:30:35.503 xcu_b1 : temps=151.0828,275.6105,143.0807,134.9645,277.6646,400.0000,400.0000,400.0000,400.0000,400.0000,162.0839,162.9854 It looks like the channel are swapped on the hardware side. |
Comment by cloomis [ 12/Jun/24 ] |
We have been here before with these b2 sensors, INSTRM-2131. Along with the discussion over whether two or three were swapped. Along with a question about b4. Never followed up with the (hardware?) tests though. I remembered putting a swap in code, but that was for a pair of b2 amps in the ccdActor ( |
Comment by cloomis [ 12/Jun/24 ] |
I think I'd vote for three being wrong: (high, low, middle) vs. (middle, high, low). 143 and 151 are different enough to be distinguished. |
Comment by arnaud.lefur [ 12/Jun/24 ] |
Yes, sorry. I agree. |