[INSTRM-2147] Update boresignt in transform.py after MCS realignment Created: 18/Jan/24  Updated: 24/Jan/24

Status: Open
Project: Instrument control development
Component/s: pfs_utils
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: yuki.moritani Assignee: chyan
Resolution: Unresolved Votes: 0
Labels: EngRun
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Sprint: PreEngRun15Mar

 Description   

MCS camera assembly was re-aligned in January 2024 after an accident at the instrument exchange.  Since the spots moved on MCE camera (by ~25pix in camera y-axis), the boresight needs to be updated. (Without update, measured FF position as well as fiber home position is shifted due to distortion.)

New bore sight at zenith is (4447.20, 2835.14)