[INSTRM-2124] R2 and B2 CCD heaters not showing as active in the GUI Created: 30/Nov/23 Updated: 31/Jan/24 |
|
| Status: | Open |
| Project: | Instrument control development |
| Component/s: | None |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Normal |
| Reporter: | yuki.moritani | Assignee: | arnaud.lefur |
| Resolution: | Unresolved | Votes: | 0 |
| Labels: | near-term | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Description |
|
This is reported to Bugzilla (https://pfs.ipmu.jp/internal/bts/show_bug.cgi?id=922) Both R2 and B2 CCD heaters seem to work correctly and can be controlled, but they still appear as inactive in the SpS GUI. |
| Comments |
| Comment by arnaud.lefur [ 30/Nov/23 ] |
|
the keywords generated by xcuActor are weird, as the ccd heater bit (5th value) is set to 0 (rubin3_ics) alefur@shell-ics:~$ oneCmd.py xcu_r2 heaters ccd power=50 2023-11-29T19:52:46.052 sent xcu_r2 heaters ccd power=50 (from alefur_shell_ics) 2023-11-29T19:52:48.181 xcu_r2 i heaters=1,0,0.000,0.000,0,0,0.500,0.000 2023-11-29T19:52:48.181 xcu_r2 : (rubin3_ics) alefur@shell-ics:~$ oneCmd.py xcu_b2 heaters ccd power=50 2023-11-29T19:53:00.412 sent xcu_b2 heaters ccd power=50 (from alefur_shell_ics) 2023-11-29T19:53:00.689 xcu_b2 i heaters=0,0,0.000,0.000,0,0,0.500,0.000 compared with r1 : (rubin3_ics) alefur@shell-ics:~$ oneCmd.py xcu_r1 heaters status 2023-11-29T19:54:01.063 sent xcu_r1 heaters status (from alefur_shell_ics) 2023-11-29T19:54:01.246 xcu_r1 i heaters=0,0,0.000,0.000,1,0,0.600,0.000 |