[INSTRM-1990] PFI temperature sensor for EBOX1 and EBOX2 is mislabeled Created: 06/Jun/23  Updated: 24/Aug/23  Resolved: 24/Aug/23

Status: Done
Project: Instrument control development
Component/s: ics_pebActor
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: yuki.moritani Assignee: chyan
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File sts_pfiTemps_newEbox_20230617.png    

 Description   

By touching the temperature sensor for EBOX 1,2,3, it was found that the sensors for EBOX 1 and EBOX2 were labelled flipped.



 Comments   
Comment by yuki.moritani [ 19/Jun/23 ]

We confirmed the wiring of the temperature sensor inside the old Ebox is the same  as the new E-box. We also confirmed wiring inside the old E-box is correct for AG3 and AG4 . Looking the graph when the AG cameras were cooled down one by one (AG6 -> AG5 -> AG4 -> AG3 -> AG2 -> AG1),  the new box  also show that the value AG3 and AG4 are flipped.

Comment by chyan [ 19/Jun/23 ]

If the wiring of both E-boxes are correct and consistent. Should we correct in software side?