[INSTRM-1940] Fix cooling-when-idle loop on pfilamps pi. Created: 24/Apr/23  Updated: 04/Aug/23  Resolved: 04/Aug/23

Status: Done
Project: Instrument control development
Component/s: ics_pfilampsActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: EngRun
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

After a lamp illumination has finished, the cooling fan is turned on for 120s. If a new lamp command comes in before that time ends, the fans can be turned off unexpectedly. Fix the logic to safely cool.



 Comments   
Comment by arnaud.lefur [ 20/May/23 ]

cloomis will talk to Jim for new suggestions.

Comment by cloomis [ 11/Jul/23 ]

We think we have a fix.

Comment by cloomis [ 04/Aug/23 ]

Fixed as part of INSTRM-2007

Generated at Sat Feb 10 16:41:14 JST 2024 using Jira 8.3.4#803005-sha1:1f96e09b3c60279a408a2ae47be3c745f571388b.