[INSTRM-1937] Add pfilamps software interlock for dome shutter Created: 23/Apr/23  Updated: 29/Dec/23  Resolved: 18/Dec/23

Status: Done
Project: Instrument control development
Component/s: ics_pfilampsActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: EngRun
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Issue Links:
Blocks
is blocked by INSTRM-2053 Add dome shutter state keyword. Done

 Description   

We should disallow lamps firing if the PFI is on the telescope and the dome shutter is not closed. [I nearly just turned the quartz on too early].

The gen2Actor has the dome shutter state and can broadcast it; the pfilampsActor can use it.



 Comments   
Comment by cloomis [ 18/Dec/23 ]

Tagged 1.3.3

Comment by yuki.moritani [ 29/Dec/23 ]

(just for recording)

When we took pfilamp data, iic failed with the following message

2023-12-29 08:33:57.059Z cmdr 20 CmdrConnection.py:212 command returned pfilamps 'prepare halogen=45'
2023-12-29 08:33:57.059Z cmds 20 CommandLink.py:122 > 2 168 w subCommand=30412,2,"pfilamps prepare halogen=45",1,"dome shutter must be closed (open) to run lamps"

However, the status on the telescope side was "closed". The status didn't seem refreshed at that time; after running "gen2 status" command, the command ran OK.