[INSTRM-190] CCD read command does not finish properly Created: 11/Sep/17 Updated: 13/Jul/18 Resolved: 13/Jul/18 |
|
| Status: | Won't Fix |
| Project: | Instrument control development |
| Component/s: | None |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Normal |
| Reporter: | arnaud.lefur | Assignee: | cloomis |
| Resolution: | Won't Fix | Votes: | 0 |
| Labels: | None | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Issue Links: |
|
||||||||
| Description |
|
For some obscure reason, from time to time ccd read command doesn't finish properly and doesn't generate its keywords. 2017-09-09 13:50:43.128Z cmds 20 Actor.py:425 new cmd: read arc exptime=2.004 obstime=2017-09-09T13:50:40.590394 2017-09-09 13:50:43.173Z cmds 20 CommandLink.py:120 > 2 1900 d text="fee is: FeeControl(port=/dev/ttyS1, device=Serial<id=0x7f162ebfe8... 2017-09-09 13:50:43.173Z cmds 20 CommandLink.py:118 > 2 1900 i exposureState=reading 2017-09-09 13:50:48.509Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=0,4300 2017-09-09 13:50:50.448Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=250,4300 2017-09-09 13:50:52.444Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=500,4300 2017-09-09 13:50:54.445Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=750,4300 2017-09-09 13:50:56.392Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=1000,4300 2017-09-09 13:50:58.371Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=1250,4300 2017-09-09 13:51:00.350Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=1500,4300 2017-09-09 13:51:02.332Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=1750,4300 2017-09-09 13:51:04.311Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=2000,4300 2017-09-09 13:51:06.289Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=2250,4300 2017-09-09 13:51:08.263Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=2500,4300 2017-09-09 13:51:10.260Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=2750,4300 2017-09-09 13:51:12.242Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=3000,4300 2017-09-09 13:51:14.221Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=3250,4300 2017-09-09 13:51:16.203Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=3500,4300 2017-09-09 13:51:18.182Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=3750,4300 2017-09-09 13:51:20.163Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=4000,4300 2017-09-09 13:51:22.152Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=4250,4300 2017-09-09 13:51:22.643Z cmds 20 CommandLink.py:118 > 2 1900 i readRows=4299,4300 2017-09-09 13:51:22.793Z ccd 30 ccd.py:179 creating fits file: /data/pfs/2017-09-09/PFLA00463612.fits 2017-09-09 13:51:24.844Z cmds 20 CommandLink.py:120 > 2 1900 f text="command failed: NameError(\"global name 'serial' is not defined\"... Neverless, the fits file is generated and seems to be correct, but as a result I have to perform the ISR manually on those exposures. |
| Comments |
| Comment by cloomis [ 13/Jul/18 ] |
|
Fixed in |