[INSTRM-1895] Arrange for the ASIC-H4 SPI bus activity be run at 20mA Created: 14/Mar/23  Updated: 14/Mar/23

Status: Open
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Unresolved Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

The default ASIC->H4 SPI driver currents are 4 mA. Eddie Bergeron pointed out those could be configured run to at 20mA. That said, we do not want to add any heat at the FPA. Also, other thermo-mechanical changes have evidently fixed the problem this addresses.

We only generate traffic on that bus when the main ASIC reconfiguration command is run, which we only do when powering the system up or when doing unusual engineering (changing the number of readout channels, IRP selection, windowed readout).

So we can wrap that command by changing the SPI driver currents from 4 to 20 mA and back to 4mA.