[INSTRM-1824] Enu slit start failed on sm2 Created: 16/Dec/22  Updated: 14/Jan/23  Resolved: 14/Jan/23

Status: Done
Project: Instrument control development
Component/s: ics_enuActor
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: fmadec Assignee: arnaud.lefur
Resolution: Done Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

error using a ditherArc sequence



 
2022-12-14 11:16:35 iic i sequence=2603,None,ditheredArcs,"SM2 nir-sn1","","iic ditheredArcs pixelStep=0.5 exptime=3 duplicate=6 cam=n2 switchOn=xenon switchOff=xenon",failed,"SpsSlit(SlitStartFailed(sm2 with RuntimeError('hexapod not in ready state, going home aborted ...') in _init()))" 2022-12-14 11:16:35 iic f text="SpsSlit(SlitStartFailed(sm2 with RuntimeError('hexapod not in ready state, going home aborted ...') in _init()))


 Comments   
Comment by arnaud.lefur [ 17/Dec/22 ]

Hum, have you repeated the procedure we did last month at Subaru for sm3 ?
Eg, uploading the hexapod script for maintaining position after reboot

Comment by fmadec [ 19/Dec/22 ]

I did not but I guess you did... because the sequence right before that one worked !

Comment by arnaud.lefur [ 14/Jan/23 ]

Scripts were uploaded on the hexapod controller.
Now it's working fine.