[INSTRM-1806] n2 (aka nir-sn1) ionpumps stop themselve wihtout tempo Created: 29/Nov/22 Updated: 30/Nov/22 |
|
| Status: | Open |
| Project: | Instrument control development |
| Component/s: | ics_xcuActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Blocker |
| Reporter: | fmadec | Assignee: | cloomis |
| Resolution: | Unresolved | Votes: | 0 |
| Labels: | None | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Description |
|
2022-11-29 09:55:35.834Z cmds 20 Actor.py:517 new cmd: ionpump on 2022-11-29 09:56:09.386Z cmds 20 CommandLink.py:121 > 1 1157244 w ionPump1Errors=0x20000,ERROR,"Pump shut down by itself"
|
| Comments |
| Comment by fmadec [ 30/Nov/22 ] |
|
test with an other controller :
2022-11-30 09:27:26.490Z cmds 20 Actor.py:517 new cmd: ionpump on 2022-11-30 09:27:42.135Z cmds 20 CommandLink.py:121 > 1 20861 w ionPump1Errors=0x10000,ERROR,"Pressure limit hit" So again less than 20s , but the error code is different this time , right ? but we should not have that error so fast
|
| Comment by fmadec [ 30/Nov/22 ] |
|
an other test: we cmd only ionpump2 to start but got an error on pump 1 ...: 2022-11-30 09:39:37 cmdIn=xcu_n2 ionpump on pump2 2022-11-30 09:39:41 xcu_n2 w ionpump1=0,0,0,38,0 2022-11-30 09:39:41 xcu_n2 w ionpump1errors=0x20000,ERROR,"Pump shut down by itself" 2022-11-30 09:39:41 xcu_n2 i ionpump2=1,1600,1.8e-06,37,8e-09 2022-11-30 09:39:41 xcu_n2 i ionpump2errors=0x00000,OK,OK pressure rises (as seen before) on pump2 and then it stopped: 2022-11-30 09:44:42.834Z cmds 20 CommandLink.py:121 > 1 21371 w ionPump2Errors=0x10000,ERROR,"Pressure limit hit" |