 Description   

I though it was solved....
in a ditherFlat sequence :

2022-11-25 00:54:45 iic w subcommand=19131,0,"sps slit start cams=r3,b3",1,"SlitStartFailed(sm3 with ValueError('failed to power on=slit outlet ...') in powerSwitch())" 


 Comments   
Comment by fmadec [ 28/Nov/22 ]

any update on this Arnaud? I can't take (dither) images on SM3.

Comment by arnaud.lefur [ 29/Nov/22 ]

That's the pdu misbehaving, as they usually do, we get a lot of socket timeout. I'll increase the number of retry + timeout value.
Note that the lamp keys are still not correct in the header for dcb data, which make them hard to reduce.
Maybe, it's better to fix this first ? Before you get a brand new dataset ?

Comment by fmadec [ 29/Nov/22 ]

absolutely ! let me know when it is done!