[INSTRM-1605] when windowing, wipe the area above the window after reading that out. Created: 14/May/22  Updated: 27/May/22  Resolved: 27/May/22

Status: Done
Project: Instrument control development
Component/s: ics_ccdActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Done Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

When windowing, we currently quickly wipe the rows to the lower edge of the readout window, read the window, then stop.

We also need to wipe the area above the window, to avoid imprinting on the next exposure.



 Comments   
Comment by cloomis [ 27/May/22 ]

Merged at c3aabc1, tagged 1.8.5

[ finished and merged during 2022-11 engineering run. ]