[INSTRM-1575] Shutter ISSUES Created: 12/Apr/22 Updated: 13/Apr/22 |
|
| Status: | Open |
| Project: | Instrument control development |
| Component/s: | ics_enuActor, ics_sequencePanel |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Major |
| Reporter: | Romain Philippe Lhoussaine ben brahim | Assignee: | Unassigned |
| Resolution: | Unresolved | Votes: | 0 |
| Labels: | ICS, R3, Red, SM3, Shutter | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Attachments: |
|
| Reviewers: | fmadec |
| Description |
|
Issues occurred on January 14th, March 31th and April 11th 2022. During data acquisition with the sequence panel with b3, the red shutters enters in fail mode. The last time, power cycle does not allow the shutter to be back on operational mode. It looks also that when the sequence failed, the CTRL turn off just after : it is not happening when you are manipulating the red shutter manually on SpsGui. |
| Comments |
| Comment by arnaud.lefur [ 13/Apr/22 ] |
|
Regarding the CTRL channel turning off, there aren't any logic like this in the code. |
| Comment by Romain Philippe Lhoussaine ben brahim [ 13/Apr/22 ] |
|
You are right ! |
| Comment by arnaud.lefur [ 13/Apr/22 ] |
(conda-ics) alefur@PFS-WS2:/software/mhs/logs/actors/enu_sm3$ cat current.log |grep 'new cmd:' 2022-04-13 07:50:27.038Z cmds 20 Actor.py:507 new cmd: biasha reboot 2022-04-13 07:50:27.046Z cmds 20 Actor.py:507 new cmd: power off=ctrl biasha reboot command power cycle the CTRL channel, but again the intent there was to reboot the arduino board, which is not the issue here. |