[INSTRM-1539] Add warning/alarm on cryocooler tip temp == 400K Created: 03/Mar/22 Updated: 03/Mar/22 Resolved: 03/Mar/22 |
|
| Status: | Done |
| Project: | Instrument control development |
| Component/s: | ics_alertsActor, ics_xcuActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Task | Priority: | Normal |
| Reporter: | cloomis | Assignee: | cloomis |
| Resolution: | Done | Votes: | 0 |
| Labels: | SPS | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Description |
|
The cryocoolers have a temperature sensor at the tip. When disconnected, that reading goes to 400K, and the cooler will not start. We have only seen this at assembly, but it just happened after a warmup and we did not notice. So add a warning and alarm for regular status queries. |
| Comments |
| Comment by cloomis [ 03/Mar/22 ] |
|
Merged at 45f5414, tagged 2.0.3 |