[INSTRM-147] CCD_r1 wrong offsets ? Created: 11/Jul/17  Updated: 13/Oct/17  Resolved: 13/Oct/17

Status: Done
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: fmadec Assignee: cloomis
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File runR1-13_CCD_channels.png    

 Description   

Dear Craig,

ccd_r1 looks strange, one chip is saturated and the other is around 11000 (as before)

Fabrice



 Comments   
Comment by cloomis [ 12/Jul/17 ]

The problems with CCD1 are far beyond bad offsets. All four amps are railed at 64k. Lowering VBB did not change anything. There are plausible non-detector causes, but that is a matter for a different venue.

I'll lower priority and leave this open for the offsets, which is blocked on the ics_config work.

Comment by cloomis [ 12/Jul/17 ]

The current method of loading offsets uses image files headers, which seemed cute at the time. I'm moving them into ics_config files.

Comment by cloomis [ 13/Oct/17 ]

Hardware fixed.