[INSTRM-141] Pressure gauge gives value while it is not connected Created: 30/Jun/17  Updated: 25/May/18  Resolved: 25/May/18

Status: Done
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Normal
Reporter: fmadec Assignee: cloomis
Resolution: Done Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Attachments: PNG File run31_pressurejump.png    

 Description   

Hi Craig,

We have strange behaviour here with pressure gauge of R0 (CRA-R2).
It was working well, then during the night there was a jump in pressure.
Not much to test while under vacuum so we decided with Rudy to unplugged it but it was still returning a value (the exact same value 9.751e-5), then we plugged it back or pressure change but after a while it reached the same pressure.

see attached plot



 Comments   
Comment by cloomis [ 25/May/18 ]

Fixed with PCM firmware 20170620_01.