[INSTRM-1238] Add xcuActor support for new heater loops Created: 15/Apr/21  Updated: 24/Jan/24  Resolved: 24/Jan/24

Status: Won't Fix
Project: Instrument control development
Component/s: ics_xcu_smb, ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Won't Fix Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Story Points: 2

 Description   

The INSTRM-1208 work added more expressive loop control commands. Actually call them.



 Comments   
Comment by cloomis [ 13/Apr/23 ]

Add heater on temp=FF.FF command and a new keyword: heaterN=name,mode,tempSet,levelSet,level, where N=(1,2), name=("asic","ccd","h4"), mode=(OFF,LEVEL,TEMP,SAFETY), level is the current heater fraction, and at most one of the tempSet or levelSet setpoints are set.

Comment by cloomis [ 24/Jan/24 ]

Done in INSTRM-1152