[INSTRM-1201] test actor failed while trying to do exposure Created: 04/Mar/21 Updated: 04/Mar/21 Resolved: 04/Mar/21 |
|
| Status: | Won't Fix |
| Project: | Instrument control development |
| Component/s: | ics_testsActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Normal |
| Reporter: | fmadec | Assignee: | arnaud.lefur |
| Resolution: | Won't Fix | Votes: | 0 |
| Labels: | SPS | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Description |
|
2021-03-03T14:56:30.297 tests w text="iic bias failed: RuntimeError('Sub-command has failed.. sequence aborted..',) in processSubCmd() at /software/mhs/products/Linux64/ics_iicActor/1.2.5/python/ics/iicActor/utils/sequencing.py:266" |
| Comments |
| Comment by arnaud.lefur [ 04/Mar/21 ] |
2021-03-03 14:56:32.655Z cmdr 10 CmdrConnection.py:58 read: client.v114.tests 61 iic w text="ccd_b3 wipe failed: KeyError('ccd',) in ccd() at python/ccdActor/main.py:52"
Somewhere to look ... |
| Comment by fmadec [ 04/Mar/21 ] |
|
maybe but I don't see the point fee controller is connected, ccd_b3 status works so I am probably missing (again) something obvious but .... |
| Comment by arnaud.lefur [ 04/Mar/21 ] |
|
ccd controller is the problem. |
| Comment by fmadec [ 04/Mar/21 ] |
|
so it was a config issue (wrong fstab) so my bad nevertheless I have a terrible error handling system I think...
|
| Comment by arnaud.lefur [ 04/Mar/21 ] |
|
problem was coming from lower software layer. |