[INSTRM-1067] Add safety routine for when the cooler auto-restarts after a long time. Created: 02/Sep/20 Updated: 02/Sep/20 |
|
Status: | Open |
Project: | Instrument control development |
Component/s: | ics_xcuActor |
Affects Version/s: | None |
Fix Version/s: | None |
Type: | Task | Priority: | Normal |
Reporter: | cloomis | Assignee: | cloomis |
Resolution: | Unresolved | Votes: | 0 |
Labels: | None | ||
Remaining Estimate: | Not Specified | ||
Time Spent: | Not Specified | ||
Original Estimate: | Not Specified |
Description |
We just had a short power outage (~10s), and the cooler auto-restarted. OK. We have an XCU routine which shuts the cooler off if the reject temp goes up, but nothing for when the cooler restarts after a longer time, after the temps and pressures have gone way up. |