[INSTRM-1067] Add safety routine for when the cooler auto-restarts after a long time. Created: 02/Sep/20  Updated: 02/Sep/20

Status: Open
Project: Instrument control development
Component/s: ics_xcuActor
Affects Version/s: None
Fix Version/s: None

Type: Task Priority: Normal
Reporter: cloomis Assignee: cloomis
Resolution: Unresolved Votes: 0
Labels: None
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified


 Description   

We just had a short power outage (~10s), and the cooler auto-restarted. OK. We have an XCU routine which shuts the cooler off if the reject temp goes up, but nothing for when the cooler restarts after a longer time, after the temps and pressures have gone way up.