[INSTRM-1046] duplicate lamp-timed exposures are mis-sequenced Created: 07/Aug/20 Updated: 11/Aug/20 Resolved: 11/Aug/20 |
|
| Status: | Won't Fix |
| Project: | Instrument control development |
| Component/s: | ics_iicActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Normal |
| Reporter: | cloomis | Assignee: | cloomis |
| Resolution: | Won't Fix | Votes: | 0 |
| Labels: | None | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Description |
|
sps.TimedArc and .TimedFlat append the dcb prepare commands to the sequence .head list, but the .expose() method appends to the regular sequence list. So all the dcb prepare commands happen before any of the exposures, and later exposures have no prepared lamps. Obviously, add dcb prepare commands to the main command sequence. |
| Comments |
| Comment by cloomis [ 11/Aug/20 ] |
|
Fixed in |